Chapter 4 ION Measurement
60 HORIBA
4.8.3 ION Valency Setting
(For Customized electrode model only)
4.9 Electrode Lot No. Setting
HINT!
To delete a registered electrode lot No., tap on the right of the electrode lot No., enter
nothing, and tap .
1. Tap on the right of the ionic valence item.
The check mark is displayed on the left of the
selected item.
2. Tap or in the ionic valence setting
screen to set the ION valency.
3. Tap .
The setting applies.
To cancel the settings, tap .
1. Tap on the right of the Electrode lot item.
2. Enter the electrode lot No. in the numerical-key
screen.
Up to 8 digits can be entered.
To cancel the settings, tap .
When the Electrode model item is set to
Customize, enter the valency of the ion to be
measured.
When an electrode lot No. is entered, the lot No.
can be displayed on data printouts or recorded
in saved data.