Chapter 2 Before Measurement (Meter SET)
F-72 33
2.15 Temperature Sensor Calibration
1. Tap on the right of the Temp. calibration
item.
The Temp. calibration setting screen is dis-
played.
To cancel the operation, tap to return to the
previous screen.
2. Display the measured temperature by the
temperature sensor connected to the instrument.
Display "------", when not connecting the temper-
ature sensor.
3. Tap on the right of the temperature sensor’s
channel item.
4. Enter the temperature with the numerical screen
and tap .
Tap when do not reflect the setting.
You can perform calibration of the temperature
sensor.