Chapter 10 Specifications
94 HORIBA
10.2 Default Settings
10.2.1 Meter Default Settings
Item Selection item/Setting range
Default
values
Security
Security management
function
Enable/Disable Disable
Hold condition Hold setting mode
EXACT/NORMAL/BRIEF/
TIME/CUSTOM/OFF (Manual)
NORMAL
In selecting
"TIME"
Time setting value 2 seconds to 999 seconds 10 seconds
In selecting
"CUSTOM"
Time setting value 2 seconds to 60 seconds 10 seconds
pH variation width 0.001 pH to 0.100 pH 0.005 pH
mV variation width 0.1 mV to 60.0 mV 0.3 mV
ORP variation width 0.1 mV to 60.0 mV 0.3 mV
Ion variation width 0.1 mV to 60.0 mV 0.3 mV
Interval
memory
Interval memory function Enable/Disable Disable
Time setting value 1 second to 999 seconds 30 seconds