Chapter 3 pH Measurement
F-72 43
3.5 pH Measurement Setting
This section describes the procedures to set the conditions of pH measurement.
The setting procedures for each item are explained below.
3.6 Indicated Resolution of pH Measurement Values Setting
1. Tap the channel setting and the measurement
item in the MEAS screen to set "CH1" and "pH".
2. Tap and tap "CH1 MEAS SET".
3. pH measurement setting items are displayed.
You will see the remaining items by dragging.
4. Select items and set the conditions.
You can switch the resolution for pH
measurement values to 0.01 pH, 0.001 pH, or
Auto.
If Auto is set, the resolution is switched
automatically to 0.001 pH when the variation of
pH measurement value is 0.01 pH or less in the
latest 10 seconds, to 0.01 pH for the rest.
This is a convenient function to judge the
stability of pH measurement values.