Chapter 4 ION Measurement
56 HORIBA
4.6 Temperature Compensation Setting
If the temperature terminals of the instruction and electrode are not connected,
temperature setting is performed in MTC even when ATC is set.
4.6.1
Solution Temperature Entry
in MTC (Manual Temperature Compensation)
1. When selecting MTC for the TEMP setting item,
the Temperature item is displayed.
Tap on the right of the Temperature item.
2. Enter the sample temperature on the numerical-
key screen.
3. Tap .
The setting applies.
To cancel the settings, tap .
There are two types of temperature
compensation for ION measurement; Automatic
Temperature Compensation (ATC) and Manual
Temperature Compensation (MTC).
In ATC, the instrument detects the solution
temperature with the connected temperature
sensor, and performs temperature
compensation for the ION values of the standard
solutions used for calibration.
In MTC, measure the solution temperature and
enter the value in advance. The instrument
performs temperature compensation using the
entered temperature.