Etigam BRIGHT-CHEQ Biological Indicator Monitoring System
BRIGHT-CHEQ Incubator Operation Manual
12
CONFIGURING THE INCUBATOR
The BRIGHT-CHEQ incubator must be properly congured prior to use. The following options
aect the overall operation of the incubator (items not marked as optional are required and
must be congured prior to using the device):
• IncubationTime - the length of the incubation period. Any BRIGHT-CHEQ BI that has
remained purple for this duration will be considered negative by the incubator. This
setting aects all incubation cells.
• Time - the current time in 12-hour or 24-hour format.
• Date - the current date.
• DisplayConguration (optional) - The contrast of the LCD can be adjusted for optimal
viewing. Additionally, the display can be inverted such that light items appear dark and
dark items appear light.
• SoundConguration (optional) - The input conrmation sound and test completion alarm
can be disabled. Additionally, the alarm that sounds when a BRIGHT-CHEQ BI is removed
from an incubation cell can be disabled.
ADJUSTING INCUBATION TIME
Main Menu
Setup Menu
Setup Menu Page 2
1. From the Main Menu, touch the SETUP key to
enter the SetupMenu.
2. From the SetupMenu, touch the MORE key to enter
SetupMenuPage2.
3. Touch the INC TIME key to adjust the duration of the
incubation period. Using the key pad that appears
on-screen, enter the incubation period duration in
hours. Use the ARROW keys to move the cursor
position, if necessary. To save and return to the
SetupMenu, touch the ENTER key. To return to the
SetupMenu without adjusting the incubation time,
touch the CANCL button.
4. If conguration is complete, touch the BACK key to
return to the MainMenu.
Touching the BACK key from either the Setup
Menu or Setup Menu Page2 will return to the
Main Menu.
The default incubation time setting is 24 hours. Etigam
recommends a reduced incubation time of 10 hours based
on our study using the FDA reduced incubation time (RIT)
protocol.
CONFIGURATION - GLOBAL SETTINGS